Gently wash the mirror with mild hand soap using alcohol to remove greasy substances such as lipstick. Do not use abrasives. Rinse well.
Dry the mirror with a soft cloth lightly then air dry until all moisture has evaporated.
Any spot left on mirror will not come off after sterilization. If the mirror cannot be cleaned right away, keep it wet or covered with a wet towel.
Never place a mirror in an ultrasonic unit.
Keep the mirror out of contact with all other instruments. Wrap the mirror into a microfiber cloth or paper towel and place it in a sterilization pouch and seal.
Autoclave @134c°for 5 mins
Do not use chemical or cold sterilization solutions. These may leave permanent streaks or spots on the surface of the mirrors.